If the occlusion (blockage) was inside your infusion set, follow the
steps below.
1.
Press f twice to turn off and to confirm the error.
2. Disconnect or remove the infusion set from your infusion site.
3. Change the infusion set and prime it.
4. Reconnect the infusion set (tubing) at your infusion site.
5. Put your pump in RUN mode.
In rare cases, the cartridge itself can be the cause of the occlusion. If an
error E4: OCCLUSION occurs again, you must change the cartridge. Follow
the steps below:
1.
Press f twice to turn off and to confirm the error.
2. Disconnect or remove the infusion set from your infusion site.
3. Change the cartridge and prime the infusion set.
4. Reconnect the infusion set (tubing) at your infusion site.
5. Put your pump in RUN mode.
Chapter 7: Alerts and Errors ACCU-CHEK Spirit Insulin Pump