Appendixes ACCU-CHEK Spirit Insulin Pump172
IEC 60601 Compliance Electromagnetic
Immunity Test Test Level Level Environment – Guidance
Radiated RF 3 V/m 10 V/m Portable and mobile RF
communications equipment
IEC 61000-4-3 80 MHz to 80 MHz should be used no closer to
2.5 GHz 2.5 GHz any part of the ACCU-CHEK
Spirit insulin pump, includ-
(IEC 60601-2-24) ing cables, than the recom-
mended separation distance
calculated from the equation
applicable to the frequency
of the transmitter.
Recommended separation
distance:
d = 0.4
.
ǰ
Ҁ
P 80 MHz
to 800 MHz
d = 0.7
.
ǰ
Ҁ
P 800 MHz
to 2.5 GHz
where P is the maximum
output po
wer rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey (a), should be less
than the compliance level in
each frequency range (b).
Interface may occur in the
vicinity of equipment
marked with the following
symbol: