Accu-Chek insulin pump Blood Glucose Meter User Manual


 
Changing the Infusion Set
When you change the infusion set, keep the following points and warnings in mind:
Infusion sets are sterile products intended for single use only. Sterility is
guaranteed for unopened packaging up to the indicated expiration date.
Do not use sterile products if the package is damaged.
Properly tighten the infusion set to the adapter to avoid leakage. Turn the
infusion set luer-lock connector until it stops. Do not turn it any further and
do not use any auxiliary tools as this may crack the infusion set luer-lock
connector and result in leakage.
Your pump cannot detect infusion set leakage. You must inspect all parts
of your infusion set at least every three hours during the day and before
you go to sleep. If there seems to be an
y loss of insulin and all parts are
correctly tightened, immediately replace the leaking component.
Immediately check your blood glucose level, because insulin delivery has
been interrupted. Take appropriate actions according to your doctor’s or
healthcare team’s instructions.
Always disconnect your infusion set before changing a cartridge or
priming. You risk uncontrolled insulin delivery into your body. Always
follow the instructions
for the infusion set you are using.
Have the following materials ready:
Your ACCU-CHEK Spirit insulin pump
A disinfectant for your skin
A new ACCU-CHEK infusion set
Chapter 3: Using Your Pump ACCU-CHEK Spirit Insulin Pump70