Accu-Chek insulin pump Blood Glucose Meter User Manual


 
Programming a Standard Bolus
There are two ways to program a Standard Bolus:
“Quick” Standard Bolus “Scroll” Standard Bolus
Using a and s on your Menu-guided using d and f
insulin pump, with bolus increments with a and s for
determined by the user programming of the amount
The bolus amount per delivery is limited to 25 units using U100 insulin.
After you program a Standard Bolus there is a delay of 5 seconds before the
bolus delivery starts. This delay lets you cancel the bolus before delivery starts
by pressing the
a or s key. Alert A8: BOLUS CANCELLED will occur. Press
f twice to confirm and turn off the alert. See the section
“Alert
A8: BOLUS
CANCELLED” (page 153) for more information.
Programming a “Quick” Standard Bolus
The following process involves only the a and s keys. You can start with
either key. The recommended steps below are those described with the first
key appearing without parenthesis, such as
a. For the alternative step press
the key in the parenthesis, such as (
s).
Note:
Because your pump beeps and vibrates to guide you safely through the
programming, you may program the “Quick” STANDARD Bolus using
one finger. After the bolus is programmed, you can count the beeps and
vibrations of your pump to verify the correct programming.
The bolus increment for the “Quick” Standard Bolus is initially set to
0.5 units.
Chapter 4: Boluses and TBRs ACCU-CHEK Spirit Insulin Pump
77