Programming a MultiWave Bolus
The MultiWave Bolus function is available in the ADVANCED user menu. It is
designed to better simulate a body’s insulin delivery. It combines an immediate
bolus delivery followed by an Extended Bolus delivery. This function can be
helpful when you have meals that include both rapid and slowly absorbed
carbohydrates.
The duration of bolus delivery can be programmed in 15-minute intervals for
up to 24 hours.
Note:
• The MultiWave Bolus is available in the ADVANCED user menu.
• You can add a Standard Bolus to an ongoing MultiWave Bolus. If you
cancel the Standard Bolus, the MultiWave Bolus remains running.
• During the delivery of a MultiWave Bolus, the programming of a
second MultiWave Bolus or an Extended Bolus is blocked. If you
need a different MultiWave Bolus, put your pump in
STOP mode
to cancel the current bolus and program a new one.
Chapter 4: Boluses and TBRs ACCU-CHEK Spirit Insulin Pump88