Accu-Chek insulin pump Blood Glucose Meter User Manual


 
Chapter 3: Using Your Pump ACCU-CHEK Spirit Insulin Pump74
5.
Hold your ACCU-CHEK Spirit insulin pump
with the adapter pointing downwards.
6. Remove the adapter and cartridge from your pump by unscrewing the
adapter from the pump casing. Pulling the adapter from the pump casing
without completely unscrewing it may result in the plunger staying attached
to the piston rod. If this should happen, please call Pump Support at
1-800-688-4578.
7. Remove the cartridge from the adapter and dispose of the cartridge properly.
8. Hold the adapter up to the light to check for signs of wear or dirt (especially
on the seals inside and outside of the adapter). It is recommended that you
replace the ada
pter a
t least every 10
th
insulin cartridge change.
9. If needed, clean the adapter using water and dry it. If it shows signs of
wear or dirt,
replace it immedia
tely.
10. Install a new cartridge and a new infusion set. See the following sections
for instructions: “Changing the Infusion Set” (page 70), “Preparing the
Cartridge” (page 42), “Connecting the Cartridge, Adapter, and Infusion Set”
(page 47), “Inserting the Cartridge” (page 49), “Priming the Infusion Set”
(page 53), “Preparing the Infusion Site and Inserting the Infusion Set”
(page 56), and “Starting Insulin Delivery” (page 65).
Note:
Check the amount of insulin remaining in the cartridge at least once
a day. Press the
f key from the RUN or STOP screen to display
the remaining cartridge content in units. Before going to sleep, make
sure that:
the cartridge contains enough insulin to last through the night,
and
that the time and date are set correctly.