Accu-Chek insulin pump Blood Glucose Meter User Manual


 
4. Move to the PUMP TIMER screen.
The days until the pump timer expires are shown.
When the PUMP TIMER reaches zero, the pump will change to
STOP mode and
no longer operate. An error, E5: END OF OPERATION, will appear on the display.
Contact Pump Support at 1-800-688-4578 to discuss your options for further
use of your ACCU-CHEK Spirit insulin pump.
See the section “Review the Time Remaining” (page 112) for more information.
Alert A6: TBR C
ANCELLED (Temporary Basal Rate Cancelled)
A temporary increase or decrease of the basal rate has been cancelled.
1. Press f twice to turn off and to confirm the alert.
2. If necessary, put your pump in RUN mode.
3. Make sure that the cancellation was intended and then program a new
Temporary Basal Rate if needed.
Note:
The alert A5: PUMP TIMER ALERT and the error E5: END OF OPERATION
are based on running time (the total time the pump is in RUN mode)
and not calendar years. However, the warranty is based on the date the
pump was purchased.
Chapter 7: Alerts and Errors ACCU-CHEK Spirit Insulin Pump152