5. Press d to move to the IMMEDIATE BOLUS screen.
Press
a or s to correct and set the immediate
bolus amount.
6. Press d to move to the BOLUS DURATION screen.
Press
a or s to correct and set the Extended
Bolus duration in increments of 15 minutes for up
to 24 hours.
7. Check the total and immediate bolus amounts and the extended bolus duration
programmed on the display
. The set values can be viewed in the BOLUS
AMOUNT, IMMEDIATE BOLUS, and BOLUS DURATION screens.
8. Press f to confirm both bolus amounts and the bolus dura
tion.
The MultiWave Bolus symbol ( ) blinks for 5 seconds (bolus delivery
start delay).
The pump beeps three times and the immediate bolus delivery begins. The
countdown of the remaining bolus continues to appear on the display.
Note:
Press d to loop between programming the total bolus amount,
immediate bolus amount, and Extended Bolus duration.
Chapter 4: Boluses and TBRs ACCU-CHEK Spirit Insulin Pump90