Accu-Chek insulin pump Blood Glucose Meter User Manual


 
Flow (delivery rate) 0.2 U/sec during infusion set
priming and bolus when using
U100 insulin.
Infusion sets ACCU-CHEK infusion sets with
luer-lock connector.
IPX 8 Protected against the effects of
temporary immersion in water
under standardized conditions
[up to 60 minutes and 8 feet
(2.5 meters)].
Maximum time for an error Plastic cartridges*
E4: OCCLUSION • at med. basal rate 1.0 U/h using
U100 insulin
5h
at min. basal rate 0.1 U/h using
U100 insulin
50h
Maximum volume* before error U100 insulin
3.5 U
Maximum pressure
400kP
a (4.0 bar)
Maximum quantity delivered at U100 insulin
1.0 U
a single fault condition
Power supply One 1.5 volt AA Alkaline battery
or one rechargeable Nickel Metal
Hydride (NiMH) AA battery. Alkaline
batteries should have a minimum
capacity of 2500 mAh and NiMH
AA rechargeable batteries a
minimum capacity of 1500 mAh.
Use only a battery charger
officially recommended by the
battery manufacturer.
Pump casing Shock, scratch, pharmaceutical
resistant plastic. All edges are
rounded.
*Determined using measurement methods according to IEC 60601-2-24-1998.
Appendixes ACCU-CHEK Spirit Insulin Pump
167