Accu-Chek insulin pump Blood Glucose Meter User Manual


 
Chapter 6: Care and Maintenance ACCU-CHEK Spirit Insulin Pump144
Therapy-Related Issues
Many factors, such as alcohol consumption, medications other than insulin,
ineffective or expired insulin, decreased activity, illness, and stress can affect
blood glucose levels. Contact your doctor or healthcare team for advice on how
to manage these or any other therapy related issues.
Troubleshooting Table
Problem
Individual characters, numbers, or
symbols on the display are shown
incompletely or not at all.
Your insulin pump does not beep
or vibrate.
Self-test incomplete (after an error
E7: ELECTRONIC ERROR).
Incorrect insulin delivery due to
incorrect time setting.
Incorrect insulin delivery due to
incorrect basal ra
te programming.
Recommended Actions
Contact Pump Support at
1-800-688-4578.
Check that the beep volume and alarm
signals are turned on. Check and
change the battery. If your insulin pump
still does not beep or vibrate, contact
Pump Support at 1-800-688-4578.
Remove the battery, wait a few
seconds, and insert it again. If the
self-test remains incomplete, contact
Pump Support at 1-800-688-4578.
The correct hourly basal rate cannot be
delivered if the time is not set correctly.
Check and reset the time if it is wrong.
Do the same check when you change
the battery.
Always review changes made in basal
ra
te programming. Confirm that you are
using your correct basal rate profile.