WFM 601A, WFM 601E & WFM 601M User Manual
1–5
Installation
This section provides instructions for installing the waveform monitor into a
standard rack or one of the optional cases. At installation time, save the shipping
carton and packing materials (including the anti-static bag) in case you need to
ship the instrument.
Included Accessories
The waveform monitor is shipped with several accessory items which may be
required for normal use. For more information on accessories, refer to page 1–2.
Hardware Installation
Because operating environments vary, the waveform monitor is not shipped with
a cabinet unless you have ordered one. All qualification testing for the waveform
monitor was performed in a 1700F00 cabinet. To guarantee compliance with
specifications, you should operate the waveform monitor in one of the cabinets
described here.
The cabinets offered for the waveform monitor provide EMI shielding, protect
against electrical shock, and protect against the accumulation of dust. Figure 1–1
shows the plain cabinet, option 1700F00. A rear panel fan supplies filtered,
cooling air which exits through the cabinet vent holes. Restricting the air flow
through the vents or the rear fan can lead to an excessive internal temperature.
NOTE. To meet EMI emission specifications, the waveform monitor must be
installed in a Tektronix 1700F00, 1700F02, or 1700F05 enclosure. The enclosure
front edges must securely contact the conductive front bezel on all four sides.
Cabinets