SECTION 11—VENTILATOR TRAY
Part No 1122170 73 Formula™ Powered Seating
SECTION 11—VENTILATOR TRAY
ƽ WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all attaching
hardware is tightened securely - otherwise injury or damage may occur.
Before adjusting, repairing or servicing the seating system, ALWAYS turn the
wheelchair power OFF, otherwise, injury or damage may result.
Pinch points exist between seat and base frames. Use caution, otherwise injury may
occur.
Preparing the Ventilator Tray for Use
ƽ WARNING
Maximum weight capacity for the vent tray is 40 lbs ± 20%. DO NOT exceed weight
capacity for vent tray - otherwise, injury or damage may occur.
CAUTION
ALWAYS disconnect the ventilator battery harness from the ventilator battery
harness extension before connecting the ventilator battery harness to the
ventilator. Otherwise damage to the product may occur.
This ventilator tray was designed to hold a ventilator that is approximately 13
inches long, 14½ inches wide, and 14 inches high. Use of ventilators larger than the
above specifications may result in damage to the ventilator or stability issues.
Wheelchairs with Recline System Only - Any ventilator installation MUST provide
adequate clearances to allow unobstructed mechanical operation of the seating
system. In addition, some ventilators may have additional clearance requirements.
ALWAYS refer to the ventilator’s instructions prior to installation for any
additional requirements.
NOTE: For this procedure, refer to FIGURE 1 on page 74.
1. Position straps to the outside of the ventilator tray.
2. Position the ventilator (not shown) on the ventilator tray.
3. Disconnect the ventilator battery harness connector from the ventilator battery
harness extension (Detail “A”).
4. Connect the ventilator battery harness to the ventilator. Refer to the ventilator
manufacturer’s installation instruction.
5. Secure strap around the ventilator and clip together.
6. Securely tighten the straps around the ventilator by pulling the ends of the strap
through the rear portion of each buckle (Detail “B”).