Invacare Powered Seating Mobility Aid User Manual


 
SECTION 6—FRONT RIGGINGS
Formula™ Powered Seating 38 Part No 1122170
NOTE: Make sure spacers are positioned properly when reassembling to prevent damage to the
frame mounting tubes.
6. Reinstall the footrest(s) onto the wheelchair. Refer to Installing/Removing Footrests on
page 35.
7. Reinstall any accessory onto the footrest(s).
Raising/Lowering Elevating Front Riggings
NOTE: For this procedure, refer to FIGURE 6.6.
1. Perform one of the following:
Raising - Pull back on the release
lever and raise front rigging to the
desired height.
Lowering - Support front rigging
with one hand away from the
release lever. Push release lever
downward with other hand.
FIGURE 6.6 Raising/Lowering Elevating
Front Riggings
Adjusting/Replacing Telescoping Front Rigging Support
ƽ WARNING
If the telescoping tubes need to be extended greater than two inches, then the seat
must be repositioned rearward to ensure stability - otherwise personal injury and/or
damage to the wheelchair and surrounding property may result.
NOTE: For this procedure, refer to FIGURE 6.7 on page 39.
1. Remove the two cap screws, spacers and threaded blocks securing the telescoping
front tube to the side rail.
2. Perform one of the following:
Slide existing telescoping front rigging support to one of six depth positions.
Remove existing telescoping front rigging.
3. Secure the telescoping front tube to the side rail at the desired depth with the existing
two cap screws, spacers and threaded blocks.
4. Repeat STEPS 1 to 3 on the opposite side, if desired.
Release
Lever