Invacare 1106628 Wheelchair User Manual


 
Part No. 1106628 47 Solara jr.
This Procedure Includes the Following:
Engaging Tilt-in-Space/Recliner Operation
Adjusting Tilt Trigger Release Cables
Positioning Tilt Locking Collars
Replacing Back/Headrest Upholstery - Recliner Models ONLY
Adjusting the Back Height - Recliner Models ONLY
WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all
attaching hardware is tightened securely. Otherwise, injury may result.
The back height, seat depth, back angle, seating system, tilt angle, seat height, size/
position of the rear wheels, size/position of the front casters, as well as the user condi-
tion directly relate to the stability of the wheelchair. Any change to one (1) or any
combination of the eleven (11) may cause the wheelchair to decrease in stability. These
adjustments MUST be performed by a qualified technician.
CAUTION
DO NOT over tighten hardware attaching to the frame. This could cause dam-
age to the frame tubing.
ENGAGING TILT IN SPACE/RECLINER OPERATION
WARNING
Make sure the occupant of the wheelchair is properly positioned in the wheelchair
BEFORE using the tilt-in-space, reclining or inclining (reverse recline) to maintain
maximum stability and safety. Refer to PROCEDURE 1 - GENERAL GUIDE-
LINES in this manual.
The Solara jr. wheelchair MUST be operated by an Assistant when in any tilt or
reclined position.
Always engage both wheel locks while using the tilt-in-Space, reclining or inclin-
ing (reverse recline) the wheelchair.
ALWAYS wear your seat positioning strap. Inasmuch as the SEAT POSITION-
ING STRAP is an option on this wheelchair (You may order with or without the
Seat Positioning Strap), Invacare strongly recommends ordering the Seat
Positioning Strap as an additional safeguard for the wheelchair user.
PROCEDURE 9TILT/RECLINE
A
N
T
I
T
I
P
P
E
R
S
TILT/RECLINE