A&D AD-4405 Scale User Manual


 
AD-4405 Weighing Indicator Page 55
15.3.1. Setting the Interval Time
Step 1 By pressing the SETPOINT(3) key while pressing the PRINT key, intval will
be displayed for 2 seconds.
Step 2 Select which interval time to set using the following keys. The item is displayed.
MODE key ....... The key to select serial output int 5i or printer int p to set.
ESC key .......... The key to return to the weight display.
ENTER key...... The key to proceed to the time setting (Step 3).
Step 3 Set the interval time using the following keys. The key operation is the same as the
time settings.
0 - 9 key.......... The key to set the value at the blinking digit, then the blinking
digit shifts to the right. (Illegal numbers are ignored) .
+/- key............. The key to shift the blinking digit to the right.
CLEAR key ...... The key to reset the settings and display the interval time stored
in memory.
ESC key .......... The key to return to Step 2 without changing the interval time
settings.
ENTER key...... The key to store the interval time then return to Step 2.
15.3.2. Interval Data Output / Print output
Set the data output mode or print mode to Manual mode respectively to interval
data output or print out.
By pressing the START(7) key while pressing the PRINT key the interval timer
starts and the first data is output or printed.
At the start, the internal printer prints the interval time setting first.
By pressing the STOP (8) key while pressing the PRINT key the interval timer
stops and the data at that time is output or printed. The internal printer prints Intv.
Stop before printing the data.
The internal printer prints unstable data in the interval print mode.
The internal printer replaces over-range data with spaces and decimal point.
15. Common Printer and Data Output