A&D AD-4405 Scale User Manual


 
AD-4405 Weighing Indicator Page 27
Data Output
Item
Parameter
Description
** 0 No data output
1
Analog output Set f31, 32, 33
2
Serial output
f30
Data output
3
Serial output (Zero suppressing)
RS-232C,
RS-422/485
*: Initial settings.
Analog Output
Item
Parameter
Description
* 0 Displayed data
1
Gross data
f31
Output data
2
Net data
f32
Weight value at 4mA output
-999999 to 999999 (Initial setting is 0)
f33
Weight value at 20mA output
-999999 to 999999 (Initial setting is
10000)
Polarity is
changed by
pressing the +/-
key.
*: Initial settings.
Serial Data Format
Item
Parameter
Description
x 0 Terminator
x 1 Device ID (selected at f06)
x 2 Code number
x 3 Data number *
x 4 Result of comparison
x 5 Date
x 6 Time
x 7 Accumulated value
x 8 Accumulation count
x 9 Stable / Over
x a Displayed weight
x b Gross weight
x C Net weight
x d Tare weight
x e Weight type (G /N / T, refer to Cf06)
f34 [Type3]
Serial data format 1
Initial value
19, 2e, 3a 4f e50
f35 [Type3]
Serial data format 2
Initial value
17 2f e30
x f Weight unit (cf. Cf07)
ZERO key :
order of output
(1-999)
Exxx indicates
the end of data
to output
NET/GROSS key:
Expand or
shorten output
data length
0 - 9 , A - F key
output data
1 The comma of the parameter display indicates that the comma (f46 0, semi-colon
for f46 1) is output after the data. To turn this mark on and off, press the 9 key while
pressing NET/GROSS key.
Output of the initial settings of f34(19, 2E, 3A 4F E50) is like ST,GR,+12345.6kg.
2 The data number *(parameter 3) increments automatically with each serial data
output from 1 to 99999 (the next of 99999 is 1). The starting number can be set
using the keys.
6.2. F-Functions