A&D AD-4405 Scale User Manual


 
AD-4405 Weighing Indicator Page 25
External Input
Item
Parameter
Description
** 0
Not used (No function)
1
ZERO key
2
TARE key
3
NET/GROSS key
4
ON/OFF key
5
PRINT key, ENTER key
6
FEED key
7
Serial data output (Format 1)
8
Serial data output (Format 2)
9
Accumulation (M+)
10
Start batching
11
Stop batching
12
Over signal, Gross over and display data out when ON
13
NET weight display when shorting the terminal.
14
Accumulated data display when shorting the terminal.
15
Execute comparison when shorting the terminal.
16
Display by the second unit when shorting the terminal.
17
All keys are enabled when shorting the terminal.
(Ignores F12 setting temporary)
18
Inhibit memory code reading when shorting the terminal.
19
Start averaging to hold
f13
EXT1
Function selection of
external input
20
Release holding data
0 to 17, 19 and 20 Same as f13
f14
EXT2
18
Code Memory (BCD 1)
0 to 17, 19 and 20 Same as f13
f15
EXT3
18
Cord Memory (BCD 2)
*: Initial settings.
Accumulation
Item
Parameter
Description
10 11
Acc. no (0) / Exec (1)
20 21
Manual (0) / Auto (1)
30 31
+ Only (0) / +/- (1)
f20 [Type2]
Accumulation Mode
40 41
OK only (0) /All data (1)
Left: Sub item, select using the
ZERO key
Right: Setting, select using the
0 , 1 keys
Initial settings: 10, 21, 31, 41
0
Add data anytime
** 1 Above 5 d
2
Above 10 d
3
Above 20 d
f21
Inhibit region for
accumulation
4
Above 50 d
The selection of the inhibit region
for accumulation.
Do not set f21 0 when Auto
accumulation (f20 21).
d: division (weighing interval) of first range. *: Initial settings.