A&D AD-4405 Scale User Manual


 
Page 12 AD-4405 Weighing Indicator
Note that the ON/OFF key does not function alone. Press the ESC key while
holding the ON/OFF key to end the calibration mode, if mis-operation. After
displaying Can5el, press the ON/OFF key to end the calibration mode and
turn the indicator off.
NOTE: When displaying Caloff , press the +/- key while pressing the ON/OFF key,
instead of CAL key. However, in some cases, settings are altered to nullify
this option.
Caution The maximum display is less than or equal to 40000 divisions. This number
is calculated from the maximum capacity divided by the minimum division.
Check the accuracy of the weighing instrument periodically.
Recommended mass, use a mass heavier than 2/3 maximum capacity.
Calibrate the scale, if it is moved to another location or the environment
has changed.
It is not necessary to set the gravity acceleration correction, when calibrating
the scale with a calibration mass at the place where the scale is used.
Enter the stable weighing data while the STABLE mark is displayed. If
unstable data is used, it may cause a weighing error. Arrange the
condition using the F00 filter function.
The span calibration needs the zero calibration data. We recommend
that you perform the span calibration immediately after the zero
calibration.
If you use the dual range function of the multi-interval scale, perform the
"Range Function", "Zero Calibration" and "Span Calibration".
5.2.
5.2.5.2.
5.2.
Calibration Procedure
Calibration ProcedureCalibration Procedure
Calibration Procedure
5.2.1. Configuring a Weighing Instrument
This section explains how to set capacity, resolution, decimal point position and display
format, weighing range and unit. Perform this procedure when installing the indicator.
When Cal5et appears on the display, enter the setting mode by pressing of the
ENTER key.
Setting the range and unit.
Single Range
Select the resolution, decimal point position and format.
¯
Specify the weighing capacity.