A&D AD-4405 Scale User Manual


 
Page 26 AD-4405 Weighing Indicator
Comparator
Item
Parameter
Description
* 0 Not used
1
Upper/lower limit comparison (2 Limits setting)
2
Upper/lower limit comparison (Target and allowance)
3
Upper/lower limit comparison (Target and % allowance)
4
HH/Hi/OK/Lo/LL comparison (4 Limits setting)
5
HH/Hi/OK/Lo/LL comparison (Target and allowance)
6
HH/Hi/OK/Lo/LL comparison (Target and % allowance)
7
Check weighing 1
8
Check weighing 2
9
Check weighing 3
10
Simple batch 1
11
Simple batch 2
f22
Comparator function
12
Simple batch 3 (Loss in weight)
10 11
Includes zero band(1)
20 21
Includes minus(1)
30 31
Stable(0) /All of the time(1)
f23 (f22 1 to 6)
[Type2]
Validation of
comparison
40 41
All of the time(0)/Start, stop(1)
Initial settings are
11,21,31 and 40.
10 11
Tare when start(1)
20 21
Stop after Full by key(0)/Auto(1)
30 31
Over/Under judgment:
After dribble flow(0) /
Batch finish(1)
f23 (f22 10 to 12)
[Type2]
Sub function for
batch weighing
40 41
No function
Initial settings are
11,21,31 and 40.
* 0 Over
1
Under
f24
Relay output
selection
2
Finish / Full (f22 9,10,12)
Effective when
f22 7,9,10 or 12.
f25
Batch finish output time
0.0 to
9.9
Relay on time by 0.1second step
(00 : continuous to zero band)
Initial setting is 0.0.
f26
Zero band
-999999 to 999999 Initial setting is 0.
*: Initial settings.
Hold
Item
Parameter
Description
** 0 Not hold (Hold function is off)
1
Manual hold
2
Auto hold
f27
Hold mode
3
Manual and Auto hold
f28
Averaging time
0.0 to
9.9
Averaging time by 0.1second step
(00 : hold at start time)
Initial setting is 0.0.
*: Initial settings.