A&D AD-4405 Scale User Manual


 
AD-4405 Weighing Indicator Page 17
ESC key ..... The key to store the parameters temporarily. Proceed to the Cal 0
display.
Press and hold the ON/OFF key and press the ESC key
No parameters are changed, CanCel is displayed and the calibration
mode is finished.
Step18 Press the ON/OFF key to turn the display off.
5.3.
5.3.5.3.
5.3.
Weighing Range Function
Weighing Range FunctionWeighing Range Function
Weighing Range Function
The weighing range function can select "single range" or "dual range". Specify each
weighing interval (division) for the multi-interval instrument. Each weighing interval is
displayed according to a net value or gross value.
Caution When single range is used, performing this function is not required.
Example 1 The gross display.
Specified parameters:
First range Range = 50.00kg, division 0.02kg
Second range Range = 100.00kg (maximum capacity), division 0.1kg
Display
0kg to 50kg : The first range, division 0.02kg.
50kg to 100kg : The second range, division 0.1kg.
Division = 0.02kg
Division = 0.1kg
0kg 50kg 100kg Gross
Weighing value
Example 2 The net display using a 40kg tare value.
Specified parameters: The same parameters as example 1.
Display
-40kg to 50kg : The first range, division 0.02kg.
50kg to 60kg : The second range, division 0.1kg.
Division = 0.02kg
Division=0.1kg
-40kg 0kg 50kg 60kg Net
Tare value 90kg 100kg Gross
40kg Weighing value
5.3. Weighing Range Function