A&D AD-4405 Scale User Manual


 
Page 54 AD-4405 Weighing Indicator
Description of "Command Mode" (For Serial Interface)
Function f40 5
Operation The serial data can be output by the serial interface command only. Data
conflicts or accidental data output by mis-operation can be avoided when
using multiple instruments with the RS-422/485 interface.
Description of "Delay for continual data"
Function f38, f42
Operation This function can be used in the "Auto print mode" and "Manual print mode".
When using a non-buffered printer, set to f42 3 and f43 1 (or f38 1).
15.2.
15.2.15.2.
15.2.
Data Number
Data NumberData Number
Data Number
The data number increments automatically with each serial data output or print out
respectively from 1 to 99999 (the next of 99999 is 1). The starting number can be set
using the keys.
Setting the Serial Data Number
Press the CODE(1) key while pressing the PRINT key at the weight display, the display
changes to data r then r xxx. xxx is the next data number. Use the 0 - 9 keys to enter
the number, then press the ENTER key to store the number and return to the weight
display.
Setting the Print Data Number
Press the SAMPLE(2) key while pressing the PRINT key at the weight display, the display
changes to data p then p xxx. xxx is the next data number. Use the 0 - 9 keys to enter
the number, then press the ENTER key to store the number and return to the weight
display.
Note 1. At the number setting, the following keys are available.
+/- key Sets the number to 1.
ZERO key Sets the number to 0.
ESC key Returns to the weight display without changing the number.
2. When setting the number to 0, it is entered as 1 automatically.
15.3.
15.3.15.3.
15.3.
Interval Data Output / Print
Interval Data Output / PrintInterval Data Output / Print
Interval Data Output / Print
The AD-4405 can output and/or print data with a constant time interval using the
internal calendar/clock function. The time intervals for serial data output and for the
printer can be set individually from 1 second up to 23 hours and 59 minutes.
Caution The interval time must be set to allow enough time to finish output or
printing out.
15. Common Printer and Data Output