A&D AD-4405 Scale User Manual


 
AD-4405 Weighing Indicator Page 29
Serial Interface
Item
Parameter
Description
** 0 Stream mode, command is not acceptable
1
Manual mode, command is effective
2
Auto print mode(+), command is effective
3
Auto print mode(+/-), command is effective
4
When accumulation, automatically output
f40
Output mode
5
Command mode (output by command only)
0
No output
** 1 Manual, Fixed format
2
Auto, Fixed format
3
Manual, Format 1 (f34)
4
Auto, Format 1 (f34)
5
Manual, Format 2 (f35)
f41
Accumulated data
output at
accumulated data
display
6
Auto, Format 2 (f35)
** 0 No delay
1
0.5 second
2
1.0 second
3
1.5 seconds
f42
Delay for continual data
4
2.0 seconds
** 0 Not used
f43
Command address
1
Use (Address determined f06)
** 0 Approx. 1 second
f44
Time out
1
No limitation
** 0 CR, LF (0Dh, 0Ah)
f45
Terminator
1
CR (0Dh)
S. I/F and Internal Printer Current Loop output
D.P. Delimitter D.P. Delimitter
* 0 Dot (.) Comma (,) Dot (.) Comma (,)
1
Comma (,) Semicolon (;) Dot (.) Comma (,)
2
Dot (.) Comma (,) Comma (,) Semicolon (;)
f46
D.P. / DelimitterSerial
Interface, Internal
Printer and Current
Loop output
3
Comma (,) Semicolon (;) Comma (,) Semicolon (;)
0
600 bps
1
1200 bps
** 2 2400 bps
3
4800 bps
f47
Baud rate
4
9600 bps
** 0 Data 7bits, Even parity
1
Data 7bits, Odd parity
f48
Data bit, parity
2
Data 8bits, Non parity
*: Initial settings. bps: bit per second.