Lifescan SureStep Pro Blood Glucose Meter User Manual


 
89
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
1 800 524-7226 (USA)
1 888 353-0800 (Canada)
No Room for
New Test
Data
Bedside unit
memory full.
Transfer data to free memory
space.
No Room for
New Test
Strip Lot
There are five new
(non-validated)
test strip lots in
bedside unit list.
Transfer data and validate
new test strip lots.
Phone Line
Busy at
Workstation
1.Workstation
modem phone
line being used
by fax machine.
2.Faulty telephone
line.
1.Workstation modem’s
phone line must be a
dedicated line. Do not
share line with a fax
machine.
2.Have phone line inspected.
QC Due for
Bedside Unit
in Less than
Two Hours
Unit will be in QC
lockout if required
QC not performed
within 2 hours.
Perform required QC tests
within 2 hours.
QC Due:
QC Missed or
Failed
QC overdue or QC
test failed.
Successfully perform all
required QC tests.
Reagent Lot
Expired
Reagent (solution)
lot expired.
Enter and use new reagent
lot. Discard any remaining
expired reagent.
Remove Strip
appears after
you insert
test strip.
Not enough
sample on test
strip or test strip
inserted too
quickly.
Remove test strip. Check
confirmation dot. If white
patches or streaks appear, not
enough blood was applied for
an accurate result. Repeat test
with new test strip.
When “Insert Strip” appears,
firmly and more slowly insert
test strip until it stops.
MESSAGE
POSSIBLE
CAUSE
SOLUTION
BU.book Page 89 Monday, October 28, 2002 11:14 PM