Invacare IRC5PO2 Oxygen Equipment User Manual


 
SECTION 22—TIMING
Part No 1148070 75 Perfecto
2
Manual Tuning
NOTE:Forthisprocedure,refertoFIGURE 22.2.
1. Turnpoweroff()andunplugtheunit.
2. Removecabinet.RefertoRemovingCabinetonpage 23.
3. PlugunitinbutDONOTturnon()untilinstructedtobelow.
4. LocatethetimingswitchontheP.C. board.
5. Insertasmallcommonorflatheadscrewdriverintotheadjustmentscrewofthetiming
switch.
6. Setthetimingswitchinitiallyatthenumber“7”position.
7. Turnuniton()andwait15minutesforunittostabilize.
8. Ifafter15minutesofruntimetheconcentrationisbelowspecification,changesetting
toanumberonepositionlowerthantheprevioussetting.
9. Letunitruntenminutes.Ifconcentrationislowerstill,changetheswitchtonumber
twopositionshigher.
NOTE:Youwillfindthattheconcentratorwillperformbetterateitherthehigherorlowersetting.
Thenormaladjustmentrangeisbetweentimingpositions3andA.
10. Continuetochangesettingonestepatatime(upordown)untiltheconcentrationis
thehighestvalueachievable.
11. Onceconcentrationhasreachedspecification,retimingiscomplete.
12. Reinstallcabinet.RefertoRemovingCabinetonpage 23.
FIGURE 22.2 Manual Tuning
/O
I
I
Manual Timing
Switch
P.C. Board
Control Panel