Invacare IRC5PO2 Oxygen Equipment User Manual


 
SECTION 7—COMPRESSOR
Part No 1148070 33 Perfecto
2
Replacing Capacitor
NOTE:Forthisprocedure,refertoFIGURE 7.2onpage 33.
1. Unplugtheunit.
2. Removethecabinet.RefertoRemovingCabinetonpage 23.
3. Usingneedlenosepliers,disconnectthespadeconnectorsfromthecapacitor
terminals.
4. Cutthetiewrapthatsecuresthecapacitortothebaseoftheconcentrator.
5. Removethecapacitorfromtheconcentratorbase.
6. InstallnewcapacitorbyreversingSTEPS36.
7. Reinstallthecabinet.RefertoRemovingCabinetonpage 23.
FIGURE 7.2 Replacing Capacitor
Capacitor
Capacitor Terminals (Spade
Connectors connect here)
Tie Wrap