SECTION 11—REGULATOR
Perfecto
2
48 Part No 1148070
SECTION 11—REGULATOR
Replacing Regulator
ƽ DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
NOTE:Forthisprocedure,refertoFIGURE 11.1onpage 49.
1. Unplugunit.
2. Removecabinet.RefertoRemovingCabinet
onpage 23.
NOTE:OnconcentratormodelswiththeSensO
2
unitONLY,removethe1/8inchtubingto
oxygensensorhousing.
3. Removethefourmountingscrewsthatsecureregulatortoproducttankcap.
4. RemoveregulatorfromproducttanklidensuringthattheOringsareremoved.
NOTE:Cleanmountingsurfaceofproducttankcap.EnsurethatthetwoʺOʺringsfortheNEW
regulatorareinstalledontotheports(underside)oftheNEWregulatorbeforeinstallation.
ƽ WARNING
Ensureproperinstallationofregulator.Noteflowarrowonregulator.ThisMUSTface
towardtheregulatoroutlettubingorflowwillbeinterruptedcausingsystemshutdown.
5. InstallnewregulatorontoproducttankbyreversingSTEPS3and4.
6. Torquemountingscrewsto5±1inchlbs.
NOTE:OnunitswithSensO
2
,reinstall1/8inchtubingtooxygensensorhousing.
7. Adjustregulatorpressureifnecessary.RefertoAdjustingRegulatoronpage 49.
8. Rununitandcheckforleaks.Referto
LeakTestonpage 76.
9. Reinstallcabinet.Referto
RemovingCabinetonpage 23.