Invacare IRC5PO2 Oxygen Equipment User Manual


 
SECTION 19—HOUR METER
Perfecto
2
64 Part No 1148070
SECTION 19—HOUR METER
Replacing Hour Meter
ƽ DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
NOTE:Forthisprocedure,refertoFIGURE 19.1.
1. Unplugunit.
2. Removecabinet.RefertoRemovingCabinet
onpage 23.
3. Removecontrolpanel.RefertoRemovingControlPanel
onpage 53.
4. Removeandlabelthetwohourmeterconnectorsandwiresconnectedtobackofhour
meter.
5. Spreadretainingclipsonhourmeterhousingthatsecurehourmetertocontrolpanel.
6. Removehourmeterbypushingmeterthroughfrontofcontrolpanel.
7. InstallnewhourmeterreversingSTEPS46.
8. Reinstallcontrolpanel.RefertoRemovingControlPanelonpage 53.
9. Reinstallcabinet.RefertoRemovingCabinetonpage 23.
FIGURE 19.1 Replacing Hour Meter
Hour Meter
Connectors
Hour Meter
Wires
Hour Meter
Housing
Retaining Clips (Hour
Meter Housing)
Back of
Control Panel
Hour Meter