Invacare IRC5PO2 Oxygen Equipment User Manual


 
SECTION 20—4-WAY VALVE/MANIFOLD ASSEMBLY
Part No 1148070 69 Perfecto
2
12. Installnewslottedpoppetintothebottomofthepilotvalvestemwithtaperedend
facingtothebottom.
13. Installpilotvalvestemandnewslottedpoppetwithcoil(withyokeandlocknut
intact)intomanifoldopening.
14. Useaflatbladescrewdriver,andtightenpilotvalvestemclockwiseuntilsnug.Torque
to25inchlbs.±5inchlbs.
15. Installnewlabelontopilotvalvecoilyoke.
16. Positionpilotvalvecoilwithyokeatanapproximate90°angletotherightasviewed
fromthefrontofvalve.
17. Usetheflatbladescrewdriverandholdthepilotvalvesteminplace.
18. Tightenlocknutclockwisewiththe9/16inchwrenchuntilsnugtosecurethestemand
coilwithyoke.DONOTovertighten.Torqueto20inchlbs±3inchlbs.
CAUTION
DONOTovertightenlocknutwheninstallingitontopilotvalvestem.Damagetothe
plasticmanifoldmayoccur.
19. PlugpowercordinandturnconcentratorON(I)toensureproperoperation.
20. Reinstallcabinet.RefertoRemovingCabinetonpage 23.
21. Ifanydifficultiesareencounteredinthisinstallation,contactInvacareTechnical
Serviceat(800)8324707.