SECTION 7—COMPRESSOR
Part No 1148070 39 Perfecto
2
CAUTION
DONOTlubricateoruseoilonanymovingparts.Thecompressoreccentricusesa
precisionsealedbearing.
DONOTcrimpthepistoncupwhenyoureplacethesleeve.Ifthecupiscrimped,itMUST
bereplaced.
3. Withbearingsideofconnectingrod/eccentricassemblyfacingthecenter(motor)of
thecompressor,slidetheassemblyontotheshaftbringingitflushtothemotor
bearing.
NOTE:Makesuretheeccentricsetscrewispositionedovertheflatoftheshaft.
4. Usingtheaccesshole,tightensetscrewto
40inchlbs(Detail“A” ofFIGURE 7.4).
5. Holdthesleevedowntothehousingwithonehandandslowlyrotatetheeccentric
withtheotherhandtoensureallthecomponentsarealignedproperly.
Fan
NOTE:Forthisprocedure,refertoFIGURE 7.4onpage 36.
1. Reinstallthefanontotheshaftinoriginalorientation.Secureintopositionusinga
rubbermallet.
Valve Plate
NOTE:Forthisprocedure,refertoFIGURE 7.3onpage 35.
1. Withthesleevelocatedandfirmlyseatedonthehousing,replacethevalveplateinthe
samemannerasitwas.
2. Repeatstep1forothervalveplate.
NOTE:MakesurethetopedgeofthesleevelocatesintheOringgrooveinthebottomofthevalve
plate.
CAUTION
MakesurethattheOringisnottwistedwhenseatedinthegrooveofthebottomofthe
valveplate.
Compressor Heads
NOTE:Forthisprocedure,refertoFIGURE 7.3onpage 35.
1. Placetheheadsontopofthevalveplates.
2. Tightentheheadscrewsinacrisscrosspatternto40inchlbs.