SECTION 6—PREVENTIVE MAINTENANCE
Perfecto
2
24 Part No 1148070
SECTION 6—PREVENTIVE
MAINTENANCE
ƽ DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
NOTE:ThePerfecto
2
Concentratorsarespecificallydesignedtominimizeroutinepreventive
maintenanceatintervalsofonceperyear.Inplaceswithhighdustorsootlevels,maintenancemay
needtobeperformedmoreoften.ThefollowingMUSTbeperformedataminimumofoneyearin
servicetoassureyearsofadditionalreliability.RefertoTroubleshootingonpage 16forplugged
filtersymptoms.Powershouldbedisconnectedbeforebeginningpreventivemaintenanceonthe
Invacareconcentrator.
Cleaning the Cabinet Filter
NOTE:Forthisprocedure,refertoFIGURE6.1.
NOTE:Thereisonecabinetfilterlocatedonthebackofthecabinet.
1. Removethefilterandcleanatleastonceaweekdependingonenvironmental
conditions.
NOTE:Environmentalconditionsthatmayrequiremorefrequentcleaningofthefiltersinclude
butarenotlimitedto:highdust,airpollutants,etc.
2. Cleanthecabinetfilterwithavacuumcleanerorwashinwarmsoapywaterandrinse
thoroughly.
3. Drythefilterthoroughlybeforereinstallation.
CAUTION
DONOToperatetheconcentratorwithoutthefilterinstalled.
FIGURE 6.1 Cleaning the Cabinet Filter