Invacare IRC5PO2 Oxygen Equipment User Manual


 
SECTION 20—4-WAY VALVE/MANIFOLD ASSEMBLY
Part No 1148070 65 Perfecto
2
SECTION 20— 4-WAY VALVE/
MANIFOLD ASSEMBLY
ƽ DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
CAUTION
Thisisamaintenancefreevalve.Openingthevalvewillvoidanyandallwarranties
applicabletothevalve.
Replacing 4-Way Valve and/or Manifold Assembly
NOTE:Forthisprocedure,refertoFIGURE 20.1onpage 67.
1. Unplugunit.
2. Removecabinet.RefertoRemovingCabinetonpage 23.
3. Cuttheonetiewrapthatsecuresthemanifoldexhausttubetothe“F”tube.
4. Removethemanifoldexhausttubefromthe“F”tube.
5. Cutthetwotiewrapsthatsecurethe“F”tubetothe4wayvalve/manifoldassembly.
6. Removethe“F”tubefromthe4wayvalve/manifoldassembly.
7. Usingahoseclamptool,removetheclampsthatsecurethevinyltubestothe4way
valve/manifoldassembly.
8. Removethevinyltubesfrombothsidesofthe4wayvalve/manifoldassembly.
9. Notetheorientationofthe4wayvalve/manifoldassemblybeforeremovingfor
correctreinstallation.
10. Lift4wayvalve/manifoldassemblyup
andoutofthebaseoftheconcentrator.
11. Disconnectspadeconnectorslocatedonthesideofthe4‐wayvalveassembly(Not
shown).
12. Removethefourmountingscrewsthatsecurethe4wayvalve/manifoldassembly
together(Detail“A”).
13. Replacethenecessarycomponent(s).(i.e.‐4wayvalve,manifoldorboth).
14. Looselyinstall
mountingscrewsthrough4wayvalveandintomanifoldassembly.
NOTE:Ensurethemanifoldgasketisinplaceon4wayvalvebeforeinstalling.