SECTION 22—TIMING
Part No 1148070 73 Perfecto
2
SECTION 22—TIMING
ƽ DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
Useextremecarewhenmakingadjustmentstothetimingswitch.DONOTallow
screwdriveroryourhandstocontacttheP. C. Boardwhenunitispluggedinand/oron.
Turnunitoffandunplugbeforeadjustingtimingswitch.Electricalshockispossible.
Toaccommodatethevaryingtoleranceswhenreplacingcomponents,anadjustabletimer
isusedtocontroltheshifttimeofthepressureequalization(P.E.)valve.The SensO
2
board
allowsforautomatic(microprocessorcontrolled)adjustmentsoftheP.E.shifttimes.Non
SensO
2
boardsallowformanualadjustment.
Autotuning
ƽ WARNING
DONOTpushtheautotunebuttonwhilepoweringup.Otherwise,thefactorysettings
willbeerased.
NOTE:Forthisprocedure,refertoFIGURE 22.1onpage 74.
1. Turnpoweroff()andunplugtheunit.
2. Removecabinet.RefertoRemovingCabinetonpage 23.
3. Plugtheunitinandturnthepoweron().
4. LocatetheP.C.board.
5. Waitforthemainvalveshifttoindicateoneofthefollowing:
•IftheP.C. boardisnew,thefrontpanelREDindicatorlightwillslowlyblink,
indicatingautotunehasinitiated.
•IftheP.C. boardisused
andhasperformedautotuningbefore,pressandrelease
thesmallpushbuttonontheP.C. board.ThefrontpanelREDindicatorlightwill
slowlyblink,indicatingtheunitisautotuningandautomaticallyadjustingtheP.E.
valvetiming.
6. Settheoutputflowto5.0L/min.
7. TheGREEN(onboard)diagnosticlightwillflashthecurrentvalueoftheP.E.valve
timingnumber.
8. PlacethecoverbackontheunitbutDONOTinstallthecoverscrews.Lettheunitrun
untilautotuningiscompleted.
9. Whenautotuningiscomplete,theREDfrontpanelindicatorlightwillchangefroma
slowblinkingtoaflashingofthefinalP.E.valvetimingnumber.
/O
I