SECTION 2—INSTALLATION/SEQUENCE OF OPERATION
Perfecto
2
14 Part No 1148070
Afterfiveminutes,iftheoxygenpurityexceeds85%±2%,theGREENlightwillcontinue
toilluminate.
Iftheoxygenlevelisnotabove85%±2%afterthefirstfiveminutes,thesystemwill
continuetomonitortheO
2
andwaitforamaximumof30minutesfromstartuptoreach
85%±2%beforeactivatinganalarm.Environmentalfactorssuchaslowvoltage,high
altitude,orageofthemachinewillaffectthetimerequiredtoreach85%±2%.
Iftheoxygenlevelisnotabove85%±2%withinthefirst30minutes,theoxygen
concentrationalarmsequencewillactivateandtheunitwillshutdown.
Whenoxygenconcentrationisabove85%±2%,thesensormeasuresoxygenpurityevery
10minutes.Ifareadingfallsbelow85%±2%,aYELLOWlightwillilluminate.Ifthe
oxygenpurityfallsbelow73%±3%theREDlight/Alarm/ShutDownmodewillactivate.