CAUTION
TorquesequenceandspecificationMUSTbeadheredtoorpossibledamagetothe4way
valvecouldresult.
15. Tightenmountingscrewsinthesequencedescribedbelow(Detail“B”).
16. Pretorquethemountingscrewsto10±2inchlbsusingthetorquesequence.
17. Themountingscrewscanbenowbetorquedto22±2inchlbsinthesamesequence.
18. Connectspadeconnectorslocatedonthesideofthe4‐wayvalveassembly.
19. Positionthe4wayvalve/manifoldassemblyinthebaseoftheconcentratoronthe
foamblockintheorientationnotedinSTEP 9.
20. Slidethevinyltubesontoeachsideofthemanifold/valveassembly.
21. Usingahoseclamptool,securethevinyltubestothe4wayvalve/manifoldassembly
withtheclamps.
22. Installthe“F”tubeintothe4wayvalve/manifoldassembly.
23. Securethe“F”tubetothe4wayvalve/manifoldassemblywithtwotiewraps.
24. Installthemanifoldexhausttubeintothe“F”tube.
25. Securethemanifoldexhausttubetothe“F”tubewithatiewrap.