Fluke 1529 Thermometer User Manual


 
8.4.2.6 SENSe<chn>:RESistance:RANGe <num>|MIN|MAX|DEF
Sets the resistance range setting.
SENS4:RES:RANG MIN
This command only applies to PRT/Thermistor channels. Executing this com
-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). The <num> parameter must be a number between 0 and 1. Specifying
the MIN, MAX or DEF parameter sets the resistance range setting to the mini
-
mum, maximum, or default value respectively.
This command may be password-protected. To access this command:
If the SYST:PASS:CONV? command returns a 0 (probe password-protec
-
tion is disabled), the password is not required.
If the SYST:PASS:CONV? command returns a 1 (probe password-protec
-
tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN <pass> command (see Section 9.4.11, Password Com-
mands).
8.4.2.7 SENSe<chn>:RESistance:WIRE? [MIN|MAX|DEF]
Returns the probe wire setting.
SENS1:RES:WIRE?
3
This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). Specifying the MIN parameter returns a value of 2. Specifying the
MAX parameter returns a value of 4. Specifying the DEF parameter returns a
value of 4.
8.4.2.8 SENSe<chn>:RESistance:WIRE <num>|MIN|MAX|DEF
Sets the probe wire setting.
SENS2:RES:WIRE MIN
This command only applies to PRT/Thermistor channels. Executing this com
-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). The <num> parameter must either be 2, 3, or 4. Specifying the MIN,
MAX or DEF parameter sets the wire setting to the minimum, maximum, or
default value respectively.
1529 Chub-E4 Thermometer Readout
User’s Guide
90