Fluke 1529 Thermometer User Manual


 
value for the specified scale parameter. Specifying the MIN, MAX, or DEF pa
-
rameter sets the scale parameter to the minimum, maximum, or default value
respectively.
This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <pass> command
(see Section 9.4.11, Password Commands).
8.4.5.10 CALibrate<chn>:PARameter:LINearity<n>? [MIN|MAX|DEF]
Returns the calibration linearity parameter for the specified channel.
CAL1:PAR:LIN1?
2.8
This command only applies to PRT/Thermistor channels. Executing this com
-
mand on Thermocouple channels generates a -294, "Incompatible type" error
message.
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion linearity suffix, <n>, specifies the resistance range parameter for PRTs (1)
or Thermistors (2). Note: For PRT/Thermistor channels, the value returned by
this command is always in ohms.
Specifying the MIN parameter returns a value of -9 when the linearity suffix
<n> is 1 (for PRTs) and -9000 when the linearity suffix <n> is 2 (for Thermis-
tors). Specifying the MAX parameter returns a value of 9 when the linearity
suffix <n> is 1 (for PRTs) and 9000 when the linearity suffix <n> is 2 (for
Thermistors). Specifying the DEF parameter returns a value of 0.
8.4.5.11 CALibrate<chn>: PARameter:LINearity<n> <num>|MIN|MAX|DEF
Sets the calibration linearity parameter for the specified channel.
CAL1:PAR:LIN1 5
This command only applies to PRT/Thermistor channels. Executing this com
-
mand on Thermocouple channels generates a -294, "Incompatible type" error
message.
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. For PRT/Thermistor channels, the calibra
-
tion scale suffix <n> specifies the resistance range parameter (in ohms) for
PRTs (1) or Thermistors (2). The <num> parameter specifies the value for the
specified linearity parameter. Specifying the MIN, MAX, or DEF parameter
sets the linearity parameter to the minimum, maximum, or default value
respectively.
This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <password> com
-
mand (see Section 9.4.11, Password Commands).
1529 Chub-E4 Thermometer Readout
User’s Guide
100