LOG:AUT:DEL 10
Specifying a number in the <num> parameter causes all auto log entries with
that label number (1 to 25) to be deleted. If the ALL parameter is specified or
no parameter is specified, all auto log entries are deleted.
8.4.7.2 LOGging:AUTomatic:COUNt? [MIN|MAX|DEF]
Returns the number of measurements to be stored in the auto log during an auto
log session.
LOG:AUT:COUN?
5000
Specifying the MIN parameter returns a value of 1. Specifying the MAX pa
-
rameter value returns a value of 8160. Specifying the DEF parameter returns a
value of 8160.
8.4.7.3 LOGging:AUTomatic:COUNt <num>|MIN|MAX|DEF
Sets the number of measurements to be stored to the auto log during an auto
log session.
LOG:AUT:COUN 100
The <num> parameter is the number of measurements to be stored in the auto
log during the auto log session. Specifying the MIN, MAX, or DEF parameter
sets the number of measurements to auto log to the minimum, maximum, or de-
fault respectively.
8.4.7.4 LOGging:AUTomatic:FREE?
Returns the number of auto log entries free and stored.
LOG:AUT:FREE?
8144,16
The response consists of two numbers separated by a comma. The first number
indicates the number of auto log entries still available. The second number indi
-
cates the number of entries already stored.
Note: Every time an auto log session is started, the first entry in the log for that
session consists of the data label and the date. All subsequent entries in the auto
log for that session consist of the channel, measurement value, units, and time.
Therefore, an auto log session that stores 100 measurements actually uses 101
log entries.
8.4.7.5 LOGging:AUTomatic:LABel? [MIN|MAX|DEF]
Returns the current auto log label setting.
LOG:AUT:LAB?
1529 Chub-E4 Thermometer Readout
User’s Guide
104