Fluke 1529 Thermometer User Manual


 
1994.0
8.4.8.9 UNIT:TEMPerature?
Returns the system temperature units.
UNIT:TEMP?
C
The response is either C for Celsius, F for Fahrenheit, or K for Kelvin.
8.4.8.10 UNIT:TEMPerature <unit>
Sets the system temperature units.
UNIT:TEMP C
The <unit> parameter is either C or CEL for Celsius, F or FAR for Fahrenheit,
or K or Kelvin for Kelvin. The *RST command sets the units to Celsius.
8.4.9 Communication Interface Commands
The communication interface commands are used for querying and setting se-
rial port and IEEE configuration parameters.
8.4.9.1 SYSTem:COMMunicate:SERial:BAUD? [MIN|MAX|DEF]
Returns the serial port baud rate setting.
SYST:COMM:SER:BAUD?
2400
Specifying the MIN parameter returns a value of 1200. Specifying the MAX
parameter returns a baud rate of 19200. Specifying the DEF parameter returns a
baud rate of 9600.
8.4.9.2 SYSTem:COMMunicate:SERial:BAUD <baud>|MIN|MAX|DEF
Sets the serial port baud rate setting.
SYST:COMM:SER:BAUD MIN
The <baud> parameter can be one of the following values: 1200, 2400, 4800,
9600, or 19200. If a value other than one listed is used, the instrument automat
-
ically selects the nearest acceptable value. Values greater than 19200 or less
than 1200 are ignored. Specifying the MIN, MAX or DEF parameter sets the
port baud rate setting to the minimum, maximum, or default value respectively.
8.4.9.3 SYSTem:COMMunicate:SERial:FDUPlex? [MIN|MAX|DEF]
Returns the serial port echo (duplex) setting.
SYST:COMM:SER:FDUP?
111
8 Digital Communications Interface
Commands