Fluke 1529 Thermometer User Manual


 
DISP:RES 3
The <num> parameter specifies the number of decimal places to display. This
parameter can be 0, 1, 2, 3, or 4. Specifying the MIN, MAX or DEF parameter
sets the display resolution to the minimum, maximum, or default value respec
-
tively. Specifying the AUT parameter sets the display resolution to automatic.
8.4.6.3 DISPlay:LAMP? [MIN|MAX|DEF]
Returns the lamp brightness setting.
DISP:LAMP?
0
Specifying the MIN parameter returns a value of 0. Specifying the MAX pa
-
rameter returns a value of 3. Specifying the DEF parameter returns a value of 0.
8.4.6.4 DISPlay:LAMP <bool>|<num>|MIN|MAX|DEF
Sets the lamp brightness setting.
DISP:LAMP ON
The lamp can be turned on or off by using either the <bool> parameter (ON or
OFF), the <num> parameter (0, 1, 2 or 3), or the MIN, MAX or DEF parame-
ters. Specifying the <bool> parameter to ON turns on the lamp. Specifying the
MIN, MAX, or DEF sets the lamp to the minimum, maximum, or default value
respectively.
8.4.6.5 DISPlay:DECimal:FORMat? [MIN|MAX|DEF]
Returns the current decimal format setting.
DISP:DEC:FORM?
0
A returned value of 0 indicates a period (.) is used as the decimal separator. A
returned value of 1 indicates a comma (,) is used as the decimal separator.
Specifying the MIN parameter returns a value of 0. Specifying the MAX pa
-
rameter returns a value of 1. Specifying the DEF parameter returns a value of 0.
8.4.6.6 DISPlay:DECimal:FORMat <num>|MIN|MAX|DEF
Sets the decimal format setting.
DISP:DEC:FORM 1
The <num> parameter can be either a 0 for a period (.) or a 1 for a comma (,).
Specifying the MIN, MAX or DEF parameter sets the decimal format to the
minimum, maximum, or default value respectively.
1529 Chub-E4 Thermometer Readout
User’s Guide
102