Fluke 1529 Thermometer User Manual


 
8.4.1.5 FORMat:STAMp?
Returns the extended measurement data format setting. See Section 9.4.1.4,
FETCh? for details.
FORM:STAM?
0
8.4.1.6 FORMat:STAMp <bool>
Enables extended measurement data format with the FETCh?, MEASure? And
READ? commands. On power up this is always set to off.
FORM:STAM ON
8.4.1.7 MEASure? [<chn>]
Returns the most recent measurement for the specified channel.
MEAS? 1
0.0127
This command is the same as the FETCh? [<chn>] command
8.4.1.8 READ? [<chn>]
Returns the most recent measurement for the specified channel.
READ? 2
0.0113
This command is the same as the ?FETCh [<chn>] command.
8.4.1.9 SENSe<chn>:DATA?
Returns the input value for the specified channel.
SENS3:DATA?
100.0291, 0.0000
The SENSe suffix, <chn>, specifies the channel (1 to 4). The parameter is
ohms for low range PRTs, Kohms for high range PRTs and thermistors, and
millivolts for thermocouples. The second returned parameter is always 0 for
PRTs and themistors and the reference junction temperature for thermocouples.
8.4.2 Measurement Control Commands
The measurement control commands are used to query and set the moving av
-
erage filter, the resistance range, the number of probe wires, and the sequence
timer. The measurement control commands for resistance range and the number
of probe wires are password protected.
1529 Chub-E4 Thermometer Readout
User’s Guide
88