Fluke 1529 Thermometer User Manual


 
8.4.5.12 CALibrate<chn>:PARameter:RJC? [MIN|MAX|DEF]
Returns the calibration RJC offset parmeter for the specified channel.
CAL1:PAR:RJC?
2.1
This command only applies to Thermocouple channels. Executing this com
-
mand on PRT/Thermistor channels generates a -294, "Incompatible type" error
message.
The RJC suffix, <chn>, specifies the channel (1 to 4). This command does not
apply to memory channels. Specifying the MIN parameter returns a value of
–9. Specifying the MAX parameter returns a value of 9. Specifying the DEF
parameter returns a value of 0.
8.4.5.13 CALibrate<chn>:PARameter:RJC <num>|MIN|MAX|DEF
Sets the calibration RJC offset parameter for the specified channel.
CAL3:PAR:RJC 1.6
This command only applies to Thermocouple channels. Executing this com-
mand on PRT/Thermistor channels generates a -294, "Incompatible type" error
message.
The RJC suffix, <chn>, specifies the channel (1 to 4). This command does not
apply to memory channels. The <num> parameter specifies the RJC value for
the specified channel. Specifying the MIN, MAX, or DEF parameter sets the
RJC parameter to the minimum, maximum, or default value respectively.
This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <pass> command
(see Section 9.4.11, Password Commands).
8.4.6 Display Commands
The display commands are used for querying and setting the display resolution,
lamp brightness, decimal format, window setting, and window fields.
8.4.6.1 DISPlay:RESolution? [MIN|MAX|DEF]
Returns the display resolution setting.
DISP:RES?
2
Specifying the MIN pamameter returns a value of 0. Specifying the MAX pa
-
rameter returns a value of 4. Specifying the DEF parameter returns AUT.
8.4.6.2 DISPlay:RESolution <num>|AUT|MIN|MAX|DEF
Sets the display resolution setting.
101
8 Digital Communications Interface