LOG:DEM:POIN?
21
Specifying the MAX parameter returns the total number of entries that can be
stored in the demand log.
8.4.7.19 LOGging:DEMand:PRINt [<num>|ALL [,<port>]]
Prints the demand log entries.
LOG:DEM:PRIN 23,1
DATA_23 1 22.862C 14:59:43 09-05-00
DATA_23 2 9.963 KO 14:59:44 09-05-00
DATA_23 3 23.097C 14:59:45 09-05-00
DATA_23 4 23.202C 14:59:46 09-05-00
If no parameters are specified, the ALL parameter is assumed and all demand
log entries are printed to the serial port. Specifying the <num> parameter prints
only the log entries with the specified label number to the serial port. Op-
tionally, the log entries may be printed to a specific port by specifying the
<port> parameter (see Table 10 on page 125, Port Numbers).
8.4.7.20 LOGging:DEMand:STORe
Stores the current measurement into the demand log.
LOG:DEM:STOR
If simultaneous scan mode is enabled, one reading is stored in the log for each
enabled channel.
8.4.7.21 LOGging:DEMand:VALue? <num>|MIN|MAX|DEF
Returns the specified demand log entry.
LOG:DEM:VAL? 10
DATA_25,2,9.961,KO,14,59,35,2000,9,5
The <num> parameter specifies the demandlogentrytoreturn.Specifyingei
-
ther the MIN or DEF parameter returns the first entry in the demand log. Spec
-
ifying the MAX parameter returns the last entry in the demand log. Each entry
in the demand log consist of the label, channel number, measurement value,
scale, time and date.
8.4.7.22 LOGging:LABel<n>:NAME?
Returns the name of the specified label.
LOG:LAB18:NAME?
DATA_18
1529 Chub-E4 Thermometer Readout
User’s Guide
108