The user-defined parameters for the thermocouple polynomial conversion are
the coefficients c
0
(C0) through c
6
(C6). The user can specify internal or exter
-
nal reference junction compensation (RJC; see Table 9 on page 125). If external
RJC is selected, the user enters the temperature of the RJC reference in the ref
-
erence junction temperature (RJT) parameter.
The output of the thermocouple, in millivolts, at 25°C with the reference junc
-
tion at 0°C, is mV (25°C). For example, for a type K thermocouple, mV(25°C)
is 1.0002.
7.2.2 Copy Probe
The COPY PROBE function allows you to copy a set of probe parameters from
one channel or memory location to another. The parameters are copied from the
user selected source channel or memory location into the current channel. The
COPY PROBE function can be password protected to restrict access. See Sec
-
tion 7.5.3, Password, for more information. If this function is password pro
-
tected, the user is prompted to enter the password to gain access. If an incorrect
password is entered, the display shows the message PASSWORD INCOR
-
RECT and access is denied. If the correct password is entered or if this function
is not password protected, the bottom portion of the display appears showing
the source channel or memory location, the probe serial number, and the con-
version type. The destination channel or memory location is displayed after the
source channel type is selected.
Use the LR buttons to select the source channel or memory location. Press
the Enter button to save the new setting. Press the Exit button to cancel and to
exit to the menu.
1529 Chub-E4 Thermometer Readout
User’s Guide
42
1
24.159C
3
24.142C
2
24.634C
4
24.015C
FROM CH:
PMEMO
SERIAL:
P5
TYPE:
PT100
TO CH:
CHAN1
EXIT
ENTER / MENU UNIT CONTRAST
1529
CHUB E-4 THERMOMETER READOUT
Figure 15 Copy Probe