Fluke 1529 Thermometer User Manual


 
8.4.2.1 INITiate
Provided for compatibility but has no effect.
INIT
8.4.2.2 INITiate:CONTinuous?
Provided for compatibility but has no effect.
INIT:CONT?
1
8.4.2.3 SENSe:AVERage:COUNt? [MIN|MAX|DEF]
Returns the moving average filter setting.
SENS:AVER:COUN?
4
Specifying the MIN parameter returns a value of 1, specifying the MAX pa-
rameter returns a value of 10, and specifying the DEF parameter returns a value
of 1. The *RST command sets the average count to the default (1).
8.4.2.4 SENSe:AVERage:COUNt <num>|MIN|MAX|DEF
Sets the moving average filter setting.
SENS:AVER:COUN MIN
The <num> parameter must be a number between 1 and 10. Specifying the
MIN, MAX, or DEF parameter sets the moving average filter setting to the
minimum, maximum, or default, respectively. The *RST command sets the
moving average to 1.
8.4.2.5 SENSe<chn>:RESistance:RANGe? [MIN|MAX|DEF]
Returns the resistance range setting.
SENS3:RES:RANG?
0
This command only applies to PRT/Thermistor channels. Executing this com
-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). Specifying the MIN parameter returns a value of 0. Specifying the MAX
parameter returns a value of 1. Specifying the DEF parameter returns a value of
0. The *RST command sets the resistance range to the default (DEF).
89