Emerson Process Management IM-106-910Xi Thermometer User Manual


 
Xi Advanced Electronics
6-6
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
11. Reinstall wiring for Alarm Outputs, Flame Status Input and/or SPS/IMPS
as applicable. See Figure 6-4 and Figure 6-5 for wiring diagrams. See
Figure 6-6 for I/O board positions in the Xi enclosure.
12. Connect the ribbon cable to the I/O Board. A new cable is supplied and
should be used if the old one is damaged.
13. Swing the Xi cover up in place and tighten the four screws.
14. Prior to operating the O
2
Probe and the Xi, all optional software
enhancements (previously enabled or not) must be enabled. Notify
Rosemount Analytical Inc. and reference the following part numbers to
enable the related software options:
NOTE
For enhanced software upgrades or to enable optional software features
previously used in your Xi configuration, contact Rosemount Analytical Inc. at
1-800-433-6076.
15. Recalibrate the O
2
Probe according to the applicable calibration
instructions in Section 4: Startup and Operation.
Figure 6-4. I/O Board Wiring Connections
Part Number Software Option
6A00269G01 Enhanced Software Option Upgrade, Stoichiometric Function
6A00269G02 Enhanced Software Option Upgrade, Programmable Reference Function
6A00269G03 Enhanced Software Option Upgrade, 850
°
C Process Function