Emerson Process Management IM-106-910Xi Thermometer User Manual


 
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
4-7
Xi Advanced Electronics
SYSTEM PARAMETER
DESCRIPTIONS
Among the parameters available through the Xi and 375/475 Field
Communicator menus are a number of "System Parameters". The system
parameters define variables that configure the Xi in the transmitter system.
System parameters are described in the following table.
CPU
PARAMETER
NAME UNITS PARAMETER DESCRIPTION
Y Version -- Software version number for the CPU board.
Y Build Num -- Software build number for the CPU board.
Y Comm Status -- Communication status between the CPU and I/O boards.
Y Restart Cntr -- Software restarts counter for the CPU board.
N Poll Addr -- Polling Address - Address used to identify a Field Device; changeable by the user to control.
N Serial Number -- I/O board serial number.
N Features -- Available advanced software features for the I/O board.
ON: feature supported.
OFF: feature not supported.
N EE Erase Count -- This is for nonvolatile memory diagnostic only.
Y Num Probe -- Number of probes configuration:
If set to 1, use One Probe configuration for the Main display.
If set to 2, use Two Probes configuration for the Main display.
Y Probe Sel -- This parameter is applicable only if the Num Probe configuration is set to 2. It defines which
probe is to be displayed.
Y Probe 1 Enable -- Probe 1 state. If enabled, the probe menu will be displayed otherwise it will not be displayed.
(No, Yes)
Y Probe 2 Enable -- Probe 2 state. If enabled, the probe menu will be displayed otherwise it will not be displayed.
(No, Yes)
Y LineX Probe -- Probe number for line x. (1, 2)
Y Line 1 Data -- Main display, line 1 data: 0 displays PV (O
2
), 1 displays SV (O
2
Temp)
Y Line 2 Data -- Main display, line 1 data: 0 displays PV (O
2
), 1 displays SV (O
2
Temp)
Y Line 3 DataL -- Main display, line 3 left hand side data:
0 displays PV (O
2
)
1 displays SV (O
2
Temp)
2 displays TV (Cell Imp)
3 displays 4V (Cell mV)
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
Y Line 3 DataR -- Main display, line 3 right hand side data:
0 displays PV (O
2
)
1 displays SV (O
2
Temp)
2 displays TV (Cell Imp)
3 displays 4V (Cell mV)
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
Y Line 4 DataL -- Main display, line 4 left hand side data:
0 displays PV (O
2
)
1 displays SV (O
2
Temp)
2 displays TV (Cell Imp)
3 displays 4V (Cell mV)
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
Y Line 4 DataR -- Main display, line 4 right hand side data:
0 displays PV (O
2
)
1 displays SV (O
2
Temp)
2 displays TV (Cell Imp)
3 displays 4V (Cell mV)
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
Y Revert Time min Xi display reverts to main display time. Also locks screen if password protection is enabled.
Y Language -- 0 displays English
1 displays Spanish
2 displays German
Y Contrast -- Display contrast: (120 to 200)
Y Alarms -- See section 5, Troubleshooting, ”Alarm Indications”.
Y Password -- Security Password
Y Enable Password -- Enables/disables security password protection