AUTOCALIBRATION
SETUP
If autocalibration is desired, the Xi must be used with either an SPS 4001B or
IMPS 4000. The Xi must be properly configured before autocalibration can
take place. Refer to the applicable SPS 4001B or IMPS 4000 instruction
manual for details on performing autocalibration. Refer to Section 4, Startup
and Operation for details on manual calibration procedures.
A Xi is shipped from the factory without autocalibration configured. This same
process must be performed any time a replacement I/O Board is installed.
Configuring Autocalibration with the Xi Keypad/Display
1. Press the MENU button once.
2. From the main menu, select PROBE 1.
3. From PROBE 1, select DETAILED SETUP.
4. From the DETAILED SETUP menu, select CAL SETUP.
5. From CAL SETUP, select Auto Cal.
6. Press the RIGHT arrow key to change the state from NO to YES.
Configuring Autocalibration with the Field Communicator
1. Use the 375/475 Field Communicator software to access the HART
menu.
2. From the DEVICE SETUP menu, select DETAILED SETUP.
3. From the DETAILED SETUP menu, select CAL SETUP.
4. From CAL SETUP, select Auto Cal.
5. Press the RIGHT arrow key to change the state from NO to YES.