Xi Advanced Electronics
2-2
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
SYSTEM
CONSIDERATIONS
A typical system installation for a Xi and O
2
Probe is shown in Figure 2-1.
Figure 2-1. Typical System
Installation
MECHANICAL
INSTALLATION
Xi Advanced Electronics
The Xi Advanced Electronics is available in a panel mounting, wall mounting,
or pipe mounting configuration. Refer to Figure 2-2 or Figure 2-3 for the panel,
wall, or pipe mounting details.
1. Ensure all components are available to install the Xi.
2. Select a mounting location near or removed from the O
2
Probe.
Consider the temperature limitations of the Xi (see "Specifications")
when selecting the mounting location.
3. Mount the Xi at a height convenient for viewing and operating the
interface. Approximately 5 ft (1,5 m) is recommended.
4. The keypad window on the Xi may have interior and exterior protective
membranes. Remove the protective membranes prior to use of the Xi
enclosure. Failure to remove the protective membranes may cause the
display to appear distorted. The membrane may be difficult or
impossible to remove after extended use at elevated temperatures.