Emerson 51-DO-03 Oxygen Equipment User Manual


 
MODEL DO-03/04 SECTION 5.0
SOFTWARE CONFIGURATION
5.11 SECURITY
1. The analyzer can be programmed to require a password for access to menus. There are three levels:
Level 1: A level 1 user can
1. Zero and calibrate the oxygen sensor
2. Calibrate the barometric pressure sensor
3. Change temperature compensation from automatic to manual and enter a manual compensation temperature
4. View diagnostic variables.
Level 2: A level 2 user can
1. Do everything a level 1 user can do
2. Change alarm setpoints
3. Rerange the 4-20 mA outputs
4. Manually test both outputs and all four alarm relays.
Level 3: A level 3 user has access to every menu item. Only a level 3 user can change passwords.
A person with no password can only view the main display.
2. Press any key to enter the main menu. Move the cursor to "Program" and
press Enter (F4).
3. Move the cursor to "Configure" and press Enter (F4).
4. Move the cursor to "Security" and press Enter (F4).
5. Use the and keys to move through the list of items. To enter a password,
press Edit (F4). Use the arrow keys to enter a three-digit password. Press
Save (F4) to store the value.
Lock all: Until the user enters the "lock all" password, all he can do is view
the main display. Entering the "lock all" password allows the user access to
all Level 1 functions.
Lock program: Entering the "lock program" password allows the user
access to all Level 2 functions.
Lock config: Entering the "lock config" password allows the user access to
all Level 3 functions.
The analyzer will accept a higher level security code at a lower level securi-
ty gate. For example, the analyzer will accept a level 2 password at a level 1
gate.
NOTES:
a. A code of 000 disables security for that level.
b. The security feature will not activate until after the timeout period has
passed with no key presses.
c. A hold condition will indefinitely prolong the timeout period.
d. Security will activate immediately if power is removed and then restored.
e. To recall a forgotten code, press and hold F4 for five seconds when the
security screen appears. The code for that level will appear.
Lock all : 0
Lock program: 0
Lock config: 0
Exit Edit
Noise rejection
Main sensor cal
Security
Exit Enter
39