Emerson 51-DO-03 Oxygen Equipment User Manual


 
MODEL DO-03/04 SECTION 5.0
SOFTWARE CONFIGURATION
TABLE 5-1. Program Settings List
Continued on the following page
ITEM CHOICES FACTORY SETTINGS
SETPOINTS
A. Alarms (Section 5.1)
1. Alarm 1 (low action)
a. if oxygen (ppm) -99 to 99 ppm 0 ppm
b. if oxygen (% saturation) 0 to 200% 0 %
c. if temperature -5 to 130°C 0.1°C
2. Alarm 2 (high action)
a. if oxygen (ppm) -99 to 99 ppm 20 ppm
b. if oxygen (% saturation) 0 to 200% 200 %
c. if temperature -5 to 130°C 130°C
3. Alarm 3 See alarm 2 See alarm 2
B. Outputs (Section 5.2 and 5.3)
1. Output 1 or 2: 4 mA setting
a. if oxygen (ppm) -99 to 99 ppm 0 ppm
b. if oxygen (% saturation) 0 to 200% 0 %
c. if temperature -5 to 130°C 0.1°C
2. Output 1 or 2: 20 mA setting
a. if oxygen (ppm) -99 to 99 ppm 20 ppm
b. if oxygen (% saturation) 0 to 200% 200%
c. if temperature -5 to 130°C 130°C
CONFIGURE
A. Display options (Section 5.4)
1. Measurement Oxygen, ozone, free chlorine, total chlorine Oxygen
2. Sensor (Oxygen only) Rosemount standard, Rosemount biopharm, or Rosemount standard
other steam sterilizable
3. Units (Oxygen only) ppm, ppb, % saturation ppm
4. Temperature units °C or °F °C
5. Output 1 mA or % of full scale mA
6. Output 2 mA or % of full scale mA
7. Language English, Français, Español, Deutsch, Italiano English
8. Main display left See section 5.5 Sensor current
9. Main display right See section 5.5 Output 1 current
10 Display contrast 00-99 (darkest) 50
11. Test timeout On or off On
12. Timeout value 1 to 60 min 10 min
Table 5-1 lists the default settings and the range of choices available for each setting. To reduce the chance of
error when configuring the analyzer the first time, enter settings in the order shown in the table.
21